Background

SIARE Engineering International Group s.r.l. is glad to introduce this new product, the result of 40 years of experience and investment in technological innovation that they are implementing in recent years. SIARE Engineering International Group s.r.l. has focused heavily on the innovation of materials, ergonomics, and ease of use of its equipment. All routine operations have been simplified and the operational procedures are “foolproof”, in this way there is no margin for the user to make incorrect or inadequate maneuvers. Even the maintenance procedures have been simplified and the parts subject to wear or deterioration have substantially decreased. The new Siaretron 4000 lung ventilator is equipped with a turbine and with a 12” TFT display, which is considerably different from all other previously manufactured versions of intensive care ventilators.

Identification

The Siaretron 4000 (hereinafter called ventilator) is a lung ventilator equipped with an innovative pneumatic system based on a “double turbine” with the advantage to deliver neonatal volumes and to prevent the risks of fires for the high oxygen concentrations presence. The Siaretron 4000 is a new generation equipment, designed for the treatment of acute and sub acute diseases and can be used on Adult, Paediatric and Neonatal (optional) patients. The lung ventilator provides new advanced features for operative modes management; it is equipped with different ventilation functions and thanks to its keyboard and decoder knob the user’s selection of most suitable settings are simplified. For those who have a basic knowledge on how ventilators for resuscitation work, the use of this equipment is intuitive and a brief training course on regard would be enough. The user’s interface (GUI) includes the keyboard, the decoder knob and the screen; the last one displays the ventilators settings and measured data, as well as various functions, allowing the operator an immediate evaluation of the patient conditions; moreover, it is possible to select and display the temporal trends of the pressure, flow, volume, the loops of flow/volume, pressure/volume and the oxygen consumption (L/min) . An immediate information management system, allows the clinician to set the alarms, collect data concerning the trend of the operating parameters (TREND) and the lung ventilator EVENTS log using the MENU. The same system allows the operator to set the PATIENT TYPE (Adult, Paediatric and Neonatal), load or erase the PATIENT DATA and in case of needs, load automatically the DEFAULT PARAMETERS of the lung ventilator.

Technical Specifications

General

  • Dimensions (W x H x D) of Ventilator unit and trolley: 530 x 1350 x 460 mm
  • Weight: 26 Kg
  • Storage Relative humidity: < 95% Temperature: from -25 to +70°C
  • Atmospheric pressure: 200hPa – 1200hPa
  • Electric power supply: 100 - 240Vac / 50 - 60Hz
  • External electric connections: 15 poles programming connector / RJ for O2 cell connection
  • Patient connections: Male conic connectors 22 mm / Female of 15 mm (according to EN ISO 5356-1:2015 norm)
  • Supply pressure (O2 ): Low pressure (max 15 l/min), High pressure 280 kPa - 600 kPa / 2.8 - 6 bar / 40 - 86 psi
  • IP degree of protection: IP21
  • Automatic leaks compensation: Max 60 l/min (NIV APCV, NIV PSV)
  • Breathing rate VC/VAC: From 4 to 150 bpm
  • Breathing rate V-SIMV e P-SIMV: From 1 to 60 bpm
  • O2 concentration: Adjustable from 21 to 100% with electronic integrated mixer.
  • Use destination: Ventilator for intensive care therapy; it is suitable for ventilation of Adult, Paediatric, and Neonatal patients (optional).
  • Inspiratory flow (FLOW): 190 l/min
  • Flow-by: Automatic
  • Flow sensor: Magnetic disturbance (patented), multi-usage type
  • Oxymeter Electronic: (value displayed in breathing parameters)
  • CO2 analyzer :Optional function (Sidestream or Mainstream module available)
  • Alarm types: By MENU: with limits set by the operator By DEFAULT: the operator cannot set them up